Welcome to the Biotage Flash Purification Blogs.

      What are good reversed-phase dry load options?

      July 23, 2019 at 2:40 PM / by Bob Bickler posted in dry load, reversed-phase, sfär

      0 Comments

      Using a “dry” loading technique with flash chromatography typically improves compound purity and overall separation quality compared to liquid loading. The reasons for this I have prophesized previously and include:

      Read More

      Setting the Right Flow Rate for Flash Column Chromatography

      July 10, 2019 at 3:16 PM / by Bob Bickler posted in Developments

      0 Comments

      How does flow rate impact my flash column chromatography separation?  This is the kind of question I frequently get.  After all, we all know that flow rates that are too high or too low can result in bad prep HPLC chromatography.  Well, this is not necessarily true in flash chromatography.

      In chromatography there are three inter-related variables which impact your separation and are represented on the  chromatographer’s triangle.

      Read More

      Which loading method should I use for purification by flash chromatography?

      July 10, 2019 at 3:14 PM / by Bob Bickler posted in Developments

      2 Comments

      Getting the most benefit from your crude sample purification with column chromatography or flash chromatography involves optimizing many variables.  In previous posts I have talked about selecting the best solvents, their ratios, and maximizing load based on TLC Rf data. These are all important chromatography-generated variables but now I would like to share some tips on actual technique differences and their impact on purification performance.

      In particular in this post I will focus on the benefits and drawbacks of liquid loading and dry loading.  Both have their place in liquid chromatography but when should one technique be used over another?

      Read More

      How do I determine loading capacity in reverse phase flash column chromatography?

      July 10, 2019 at 3:01 PM / by Bob Bickler posted in Developments, Flash Theory

      0 Comments

      As the popularity of prep-scale, reversed-phase flash chromatography increases, so does the frequency that I get asked this question, "How do I determine loading capacity in reversed-phase flash chromatography?"

      In the world of HPLC, loading capacity isn’t normally a concern as it is primarily an analytical technique.  In the synthetic organic chemistry world, most purification is performed with silica gel where flash column purification methods are developed and loading capacity estimated from TLC data.  However, when normal-phase flash does not work and reversed-phase flash is needed, the question of how to determine reversed-phase loading capacity comes up.

      In this post I will attempt to provide some guidelines to help you understand and determine reversed-phase flash chromatography loading capacity.

      Read More

      How do I develop a reversed-phase flash column chromatography method?

      July 10, 2019 at 2:53 PM / by Bob Bickler posted in Developments

      10 Comments

      Method transfer from reversed-phase TLC (thin layer chromatography) to reversed-phase flash column chromatography can be very challenging.  Because of this, I often recommend using HPLC for reversed-phase flash chromatography method development. This really is a straight-forward process if you start with the right HPLC column and know a little about your HPLC system's detector.

      In this post, I will share some tips on how to develop a reversed-phase flash purification method using HPLC.

      Read More

      How to minimize particulate build-up in flash chromatography systems

      July 9, 2019 at 3:23 PM / by Bob Bickler posted in blockage, pressure, particulate

      0 Comments

      In order to perform flash chromatography consistently, the equipment you use must be properly maintained by following some “best practices”. These best practices include using clean solvents (I typically use ACS grade), volatile organic modifiers, and quality columns.

      Read More

      How much DMF or DMSO can I inject on my reversed-phase flash chromatography column?

      June 13, 2019 at 5:05 PM / by Bob Bickler posted in Developments

      0 Comments

      In a previous post I shared results of experiments where I evaluated selected organic solvents for sample dissolution and injection for reversed-phase flash purification.  I demonstrated that DMF and DMSO both are excellent solvents for this purpose and actually provide better chromatography than methanol, acetonitrile, and acetone.

      In this post I report some surprising results from follow-on work evaluating the impact of increased injection volume using DMF and DMSO as the sample diluent/injection solvent.

      Read More

      How do I Choose the Right Column Size for Purification by Flash Chromatography?

      June 13, 2019 at 4:59 PM / by Bob Bickler posted in Developments, Flash Theory

      6 Comments

      In all my years of working with medicinal and organic chemists, I have found that choosing how many grams of silica to use for purification by flash chromatography is something frequently guessed at. Getting the size of the column right is awfully important because using too few grams of silica will doom your purification to failure and using more an optimal mass of the stationary phase means the purification  consumes excess silica, solvents, and a chemist's time. To determine the optimal amount of silica for a purification, I rely on a factor called ΔCV (delta Column Volume) to identify the best loading capacity on any cartridge. I have also found that ΔCV this is a better loading capacity predictor for flash purification than ΔRf.

      Read More

      Benefits of acetonitrile over methanol in normal-phase flash column chromatography

      June 13, 2019 at 4:57 PM / by Bob Bickler posted in Developments

      2 Comments

      When it comes to the purification of polar organic compounds many chemists turn to normal-phase flash chromatography with dichloromethane and methanol as the mobile phase. This solvent system often can be challenging to optimize due to methanol’s high polarity and protic chemistry.

      I have found that acetonitrile can often replace methanol as the polar modifier in DCM-based solvent systems.  In this post I will show an example where this is true.

      Read More

      How to Optimize TLC to Enhance Purification by Flash Chromatography

      June 13, 2019 at 4:53 PM / by Bob Bickler posted in Flash Theory

      6 Comments

      In this article I discuss the optimization of solvent ratios to generate ideal Rf (retention factor) values on TLC plates.  Then I show how maximizing efficiency of flash chromatography achieves higher loading with rapid and reliable isolation of compounds, reduced solvent use and improved separation.

      Read More

      How changing stationary phase chemistry can impact separation selectivity

      June 12, 2019 at 2:42 AM / by Bob Bickler posted in selectivity, stationary phase, impurity

      0 Comments


      Challenging separations, we all have faced this vexing problem. You synthesized your compound, analyzed it, and know your molecule is in there, based on LC-MS or TLC. Then, you do some method development using a silica TLC plate and see a major spot with some minor, early-eluting impurities. You think that the purification will be easy only to find that your “purified” compound has some co-eluting impurities. Now what? Should you change solvents or change stationary phase?

      In this post, I will show how changing the column media but keeping the same solvents removed a co-eluting impurity in one of my reaction mixes.

      Read More

      Pesticide remediation of Cannabis oil - Myclobutanil removal by flash chromatography

      June 11, 2019 at 3:51 PM / by Brian Wagner posted in cannabis

      6 Comments

      Welcome to a second installment of the Biotage Cannabis Application development flash blog.  The first post, dated August 20  2016, outlined an orthogonal approach to isolating cannabinoids from winterized extract.  Give it read if you have not seen it

      Today we investigate the “hot” topic of pesticide elimination from extract.  Specifically – Myclobutanil remediation.

      Read More

      How many ways can you load sample on your column?

      May 31, 2019 at 4:02 PM / by Raffaella Bombarda posted in dry load, samplet, Samplet(R), flash chromatography, Flash column chromatography, column, silica column, flash purification, sfär, loading capacity

      0 Comments

      Biotage®, a pioneer in Flash Purification, launched the unique, removable cap SNAP flash chromatography columns in 2007. This beneficial column design feature continues with the newest Biotage flash columns named Sfär columns.

      Read More

      Dry loading media options – diatomaceous earth

      May 28, 2019 at 10:47 PM / by Bob Bickler posted in dry load, diatomaceous earth

      0 Comments

      Various flash chromatography sample loading options are available including liquid and dry loading. Choosing the right technique is important because your sample loading choices (sample solvent and dry load sorbent), can have a major impact on the results.

      In this post, I compare the two techniques and show the benefits dry loading with a form of diatomaceous earth can bring to your purification.

      Read More

      How can I perform normal-phase and reversed-phase column chromatography on one flash system?

      May 20, 2019 at 2:21 PM / by Bob Bickler posted in normal phase, reversed-phase

      0 Comments

      For chemists, flash chromatography is part of their everyday synthesis workflow. For most syntheses, crude reaction mixtures are purified by normal-phase (aka adsorption) chromatography.  There are times, however, where the crude mixture’s complexity and polarity make normal-phase chromatography very challenging.  For these situations, reversed-phase (aka partition) chromatography may be a preferred option.

      But, if you have only one flash system available, can you, should you, and how do you efficiently switch from non-polar, normal-phase solvents to polar, reversed-phase solvents – and back again without issues? In this post I'll attempt to shed some light on the topic.

      Read More

      Minor Cannabinoid Separation by Reversed-phase Flash Chromatography

      May 17, 2019 at 8:30 PM / by Bob Bickler posted in hemp, purify, cannabinoid, isolate

      0 Comments

      Previously, I have posted on a normal-phase flash chromatography method to separate and isolate CBG from a CBD-rich hemp distillate. CBG is just one of many naturally occurring minor cannabinoids of interest in this fast-growing market.

      Read More